Technical specification:
Power supply : AC220V,50HZ
Power input: ≤1300VA
Air Temperature control range:25℃~37℃ can override to 39℃
Skin Temperature control range:34℃~37℃ can override to 38℃
Sensor precision: ≤0.3℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃(bed is on the level position)
Temperature rising time: ≤40min(environmental temperature +22℃)
Noise level: ≤55dB(A)(environmental noise ≤40dB)
Inclination of bassinet: 0°~5°
Adjustment height of lift stand: < 200 mm
Indication range of oxygen concentration:0%~99%
setting range of oxygen concentration:20%~60%(1% each time)
Indication precision of oxygen concentration:±3%
Humidity indication range:0%RH~90%RH
Humidity control range:20%RH~90%RH(1% each time)
Humidity indication precision:±6%RH(environmental temperature 20℃~40℃,environmental humidity 10%RH~90%RH)
Indication presicion of humidity:±6%RH
Capacity of water tank:1000 ml
Working time of water tank when it is filled with water:≥12 hours ( less than 60%RH)
Alarming:
audible and visible alarm indication, and LED display can indicate the real temperature (except for power failure alarm and system alarm), For the following alarm 1-6(except for lower deviation), temperature controller can cut off the power supply of heater automatically.
Over-temp alarm: when the air temperature control ≤38℃(setting temperature is less than 37℃ ), or≤40℃(setting temperature is more than 37℃);skin temperature control ≤40℃;
Deviation alarm: air temperature control ±3.0℃, skin temperature control ±1.0℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or skin sensor 1 is not in the right position;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.
oxygen concentration alarm:±5%O2;
humidity deviation alarm:±15%RH。
Package:
Gross Weight:Main body 70Kg / Cabinet 80Kg
Measurement:Main body 1100mm×707mm×873mm
Cabinet 1029mm×697mm×816mm
Notice:
Some technical data of this model can be changed properly in standard permission range according to the standard and user’s requirement like oxygen concentration monitoring system and humidity control system, lift stand, blue light radiant box, infant scale, tray. If the technical specification of above product changes, please refer to the standard at selling time.