Introduction:
Curny-terner configuration diffraction grating monochromator, Aspherical light source optics & Precise grid, whole aluminum metal alloy die-castings;Simple & clear keyboard operation is convenient to realize auto 0% & 100% adjustment and error-free T/A transformation;Factor setting & direct concentration readout functions;Using universal Φ12+0.4×100 mm testing tubes or cuvette to make measure directly;Spacious sample compartment, 4 position cell rack, adaptable for 1-5 cm optical path rectangular cells.
Standard configure:
S23A spectrophotometer, glass cuvette, Power cable, Operation manual, Fuses(2A)
Technical specification:
Wavelength Range: 340-950nm
Wavelength Accuracy: ±2.5 nm
Band width: 6 nm
Photometric Accuracy: ±2%(τ) (tube), ±0.5%(τ) (cuvette)
Photometric Reproducibility: 0.3%(τ)
Stray Light: ≤0.5%(τ)
Package:
Gross Weight:9Kg
Measurement:370mX320mm×195mm
Notice:
Some technical data of this model can be changed properly according to the standard and user’s requirement. If the technical specification of above product changes, please refer to the standard at selling time.