Introduction:
The emission monochromator adopts 1200 line diffraction grating, large aperture and non-spherical reflecting mirrors produces extra high sensitivity, emission wavelength is 200-650 nm,provide better fluorescence linearity and sensitivity;High stable/long life 150W xenon lamp,continuously spectrum,no ozone;High performance photomultiplier provide better S/N;Automatic zero adjustment and automatic background subtraction.
Standard configure:
F95 Spectrophotometer, quartz fluorescence cuvette 10mm, Power cable, Operation manual, Fuses(2A)
Technical specification:
Emission monochromator wavelength Accuracy: ±2nm
wavelength repeatability: 1nm
Sensitivity: S/N≥50 (Raman peak of water in 1 cm quartz fluorescence cuvette)
Linearity deviation: ≤±3.0%
Baseline shifting: ≤±0.3%( within 10min)
Upper limit of display: ≤1.5%( within 10min)
Reading display variation: ≤1.0%(at 220V±22V)
Package:
Gross Weight:14Kg
Measurement:380mX425mm×240mm
Notice:
Some technical data of this model can be changed properly according to the standard and user’s requirement. If the technical specification of above product changes, please refer to the standard at selling time.